Fluke 5500a Coil Manual

Size:Download

®

Fluke 5500a Specifications

The 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrating clamp-type current meters that operate by two different principles—as current transformers (AC only), and by the Hall Effect (Both AC.

  1. 5500A multi-product calibrator can accurately calibrate digital & analog multimeters, thermometers, wattmeters, data loggers & many other measurement tools.
  2. Connect the Fluke 5500A coil to the Fluke 5522A calibrator and clamp the UUC to the toroid coils of the Fluke 5500A coil as illustrated in the Figure. Clamp meter Current Function Calibration Setup Note that for a current above 20A, transfer the HI lead wire to the 20A – terminal.
  3. Fluke Calibration produces the broadest range of electrical, RF, temperature, pressure & flow calibration instruments & software that help measurement professionals around the globe.

Instruction Sheet

Model 5500A/COIL

50-TurnCurrent Coil

Introduction

The 5500A/COIL is a 50-TurnCurrent Coil used as a tool for calibratingclamp-typecurrent meters that operate by two differentprinciples—ascurrent transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current source calibrator one can effectively multiply the current of the current source calibrator by a factor of 50 to support the calibration and verification of theseclamp-typecurrent meters.

PWARNING

UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard.

Using the Coil

Clamp-typecurrent meters operate as current transformers, with differing degrees of magnetic coupling between primary and secondary that vary from meter to meter. The position of the clamp meter with respect to the cable also affects the magnetic coupling between primary and secondary of the current transformer, which causes variation in reading of

the current meter. This is important to understand in order to make the most accurate and repeatable measurements. The base of the 5500A/COIL was designed so the current clamp can be centered carefully on the coil, minimizing operator error for best repeatability. Calibration accuracy to specifications is guaranteed only when proper clamp alignment is made. The clamp current meter should be centered as much as possible on the base during calibration and verification. If the clamptype current meter has alignment marks, the alignment marks should align the clamp with the center bundled wire of the 50-Turn5500A/COIL. (See Figure 1).

Figure 1. Positioning the Clamp

PN 600616 August 1995 Rev.8, 1/09

1995-2009Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.

Specifications

Number of Turns

50

Maximum Current

11A rms, continuous

20.5A rms, 2 minutes

Maximum Duty Cycle Derating

< 11A, continuous

> 11A, 2 minutes ON, 8 minutes OFF

Maximum Voltage

7 V dc or 7 V rms ac

Minimum Inner Diameter of Clamp Jaws

2.54 cm [1”]

Specification for Clamp Meter/Coil interaction

Specs

Specification for 5500A/COIL

Coil Specifications [clamp/coil interaction]

Calibrator Output

Effective

'other-type'of Clamps

Current Output

Toroidal-typeClamps

[such as the Fluke 80i-KW,-400,

[such as the Fluke 80i-600and80I-1000]

80i-410,-500,-1010and30-series]

Magnitude

Frequency

Amp-turns

±(% of output + Amps)

±(% of output + Amps)

0.2

- 0.33 A

DC

10 - 16.4999

0.50

0.02

0.33 - 2.9999 A

DC

16.5 - 149.999

Not applicable

0.5

0.14

3.0

- 20.5 A

DC

150 - 1025

0.5

0.5

0.2

- 0.33 A

45

Hz to 65 Hz

10 - 16.4999

0.28

0.003

0.56

0.03

0.33 - 2.9999 A

45

Hz to 65 Hz

16.5 - 149.999

0.28

0.025

0.56

0.25

3.0

- 20.5 A

45

Hz to 65 Hz

150 - 1025

0.28

0.09

0.56

0.9

0.2

- 0.33 A

65

Hz to 440 Hz

10 - 16.4999

0.79

0.003

1.00

0.03

0.33 - 2.9999 A

65

Hz to 440 Hz

16.5 - 149.999

0.79

0.027

1.00

0.25

3.0

- 20.5** A

65

Hz to 440 Hz

150 - 1025

0.79

0.1

1.00

0.9

0.2

- 20.5 A

Above 440 Hz

10 - 1025

Usable but unspecified

**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range.

Calculating Total Specification

The total specification of the effective current that the clamp-typemeter measures is a function of the clamp/coil interaction and of the current calibrator. To determine the total specification use the following formula:

Stotal = S 2 coil+ U 2 source

Example:

Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective current of 200A, 60 Hz) and we are calibrating a toroidal-typeclamp meter. The Calibrator’s 1 year specification at 4 Amps is ± (0.06% + 2 mA), so the effective current in the coil bundle will have a specification of ± (0.06% + 0.1A). Next, we find the total specification of the calibrator and the coil as a percentage of the output:

Specification of effective calibrator current in coil bundle = ± (0.06% + 0.1A) = 0.11% Specification due to clamp Meter/Coil Interaction = ± (0.28% + 0.09A) = 0.325%

The RSS of these two specifications determines the total specification of the clamp/source combination:

Stotal =0.325%2 + 0.11%2 = 0.343%

Model 5500a/coil, Instruction sheet, Turn current coil

Advertising

PN 600616 August 1995 Rev.8, 1/09

©1995-2009 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.

Instruction Sheet

Fluke 5500a Coil Manual 2017

Model 5500A/COIL

50-Turn Current Coil

Introduction

The 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrating clamp-type current meters that
operate by two different principles—as current transformers (AC only), and by the Hall Effect (Both AC
and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. How-
ever, by using the 50 turns of the 5500A/COIL in conjunction with a current source calibrator one can ef-
fectively multiply the current of the current source calibrator by a factor of 50 to support the calibration
and verification of these clamp-type current meters.

P

WARNING

Fluke 5500a For Sale

UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard.

Using the Coil

Clamp-type current meters operate as current transformers, with differing degrees of magnetic coupling
between primary and secondary that vary from meter to meter. The position of the clamp meter with re-
spect to the cable also affects the magnetic coupling between primary and secondary of the current trans-
former, which causes variation in reading of
the current meter. This is important to under-
stand in order to make the most accurate and
repeatable measurements. The base of the
5500A/COIL was designed so the current
clamp can be centered carefully on the coil,
minimizing operator error for best repeatabil-
ity. Calibration accuracy to specifications is
guaranteed only when proper clamp alignment
is made. The clamp current meter should be
centered as much as possible on the base dur-
ing calibration and verification. If the clamp-
type current meter has alignment marks, the
alignment marks should align the clamp with
the center bundled wire of the 50-Turn
5500A/COIL. (See Figure 1).

®

Fluke 5500a Coil Price

Figure 1. Positioning the Clamp

Fluke 5500a Coil Specs

Advertising
Popular Brands

Fluke 5500a Coil Calibration

Popular manuals