The 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrating clamp-type current meters that operate by two different principles—as current transformers (AC only), and by the Hall Effect (Both AC.
- 5500A multi-product calibrator can accurately calibrate digital & analog multimeters, thermometers, wattmeters, data loggers & many other measurement tools.
- Connect the Fluke 5500A coil to the Fluke 5522A calibrator and clamp the UUC to the toroid coils of the Fluke 5500A coil as illustrated in the Figure. Clamp meter Current Function Calibration Setup Note that for a current above 20A, transfer the HI lead wire to the 20A – terminal.
- Fluke Calibration produces the broadest range of electrical, RF, temperature, pressure & flow calibration instruments & software that help measurement professionals around the globe.
Instruction Sheet
Model 5500A/COIL
Introduction
The 5500A/COIL is a
PWARNING
UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard.
Using the Coil
the current meter. This is important to understand in order to make the most accurate and repeatable measurements. The base of the 5500A/COIL was designed so the current clamp can be centered carefully on the coil, minimizing operator error for best repeatability. Calibration accuracy to specifications is guaranteed only when proper clamp alignment is made. The clamp current meter should be centered as much as possible on the base during calibration and verification. If the clamptype current meter has alignment marks, the alignment marks should align the clamp with the center bundled wire of the
Figure 1. Positioning the Clamp
PN 600616 August 1995 Rev.8, 1/09
Specifications
Number of Turns | 50 |
Maximum Current | 11A rms, continuous |
20.5A rms, 2 minutes | |
Maximum Duty Cycle Derating | < 11A, continuous |
> 11A, 2 minutes ON, 8 minutes OFF | |
Maximum Voltage | 7 V dc or 7 V rms ac |
Minimum Inner Diameter of Clamp Jaws | 2.54 cm [1”] |
Specification for Clamp Meter/Coil interaction

Specification for 5500A/COIL
Coil Specifications [clamp/coil interaction] | ||||||||||
Calibrator Output | Effective | |||||||||
Current Output | ||||||||||
[such as the Fluke | ||||||||||
[such as the Fluke | ||||||||||
Magnitude | Frequency | ±(% of output + Amps) | ±(% of output + Amps) | |||||||
0.2 | - 0.33 A | DC | 10 - 16.4999 | 0.50 | 0.02 | |||||
0.33 - 2.9999 A | DC | 16.5 - 149.999 | Not applicable | 0.5 | 0.14 | |||||
3.0 | - 20.5 A | DC | 150 - 1025 | 0.5 | 0.5 | |||||
0.2 | - 0.33 A | 45 | Hz to 65 Hz | 10 - 16.4999 | 0.28 | 0.003 | 0.56 | 0.03 | ||
0.33 - 2.9999 A | 45 | Hz to 65 Hz | 16.5 - 149.999 | 0.28 | 0.025 | 0.56 | 0.25 | |||
3.0 | - 20.5 A | 45 | Hz to 65 Hz | 150 - 1025 | 0.28 | 0.09 | 0.56 | 0.9 | ||
0.2 | - 0.33 A | 65 | Hz to 440 Hz | 10 - 16.4999 | 0.79 | 0.003 | 1.00 | 0.03 | ||
0.33 - 2.9999 A | 65 | Hz to 440 Hz | 16.5 - 149.999 | 0.79 | 0.027 | 1.00 | 0.25 | |||
3.0 | - 20.5** A | 65 | Hz to 440 Hz | 150 - 1025 | 0.79 | 0.1 | 1.00 | 0.9 | ||
0.2 | - 20.5 A | Above 440 Hz | 10 - 1025 | Usable but unspecified |
**Note: Some calibrators may not have adequate compliance voltage to drive coil in this range.
Calculating Total Specification
The total specification of the effective current that the
Stotal = S 2 coil+ U 2 source
Example:
Assume we are driving the coil with the Fluke 5500A at 4A, 60 Hz (the clamp meter will see an effective current of 200A, 60 Hz) and we are calibrating a
Specification of effective calibrator current in coil bundle = ± (0.06% + 0.1A) = 0.11% Specification due to clamp Meter/Coil Interaction = ± (0.28% + 0.09A) = 0.325%
The RSS of these two specifications determines the total specification of the clamp/source combination:
Stotal =0.325%2 + 0.11%2 = 0.343%
Model 5500a/coil, Instruction sheet, Turn current coil
PN 600616 August 1995 Rev.8, 1/09
©1995-2009 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
Instruction Sheet
Fluke 5500a Coil Manual 2017
Model 5500A/COIL
50-Turn Current Coil
Introduction
The 5500A/COIL is a 50-Turn Current Coil used as a tool for calibrating clamp-type current meters that
operate by two different principles—as current transformers (AC only), and by the Hall Effect (Both AC
and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. How-
ever, by using the 50 turns of the 5500A/COIL in conjunction with a current source calibrator one can ef-
fectively multiply the current of the current source calibrator by a factor of 50 to support the calibration
and verification of these clamp-type current meters.
P
WARNING
Fluke 5500a For Sale
UNFUSED. Exceeding current and voltage ratings can cause burn or fire hazard.
Using the Coil
Clamp-type current meters operate as current transformers, with differing degrees of magnetic coupling
between primary and secondary that vary from meter to meter. The position of the clamp meter with re-
spect to the cable also affects the magnetic coupling between primary and secondary of the current trans-
former, which causes variation in reading of
the current meter. This is important to under-
stand in order to make the most accurate and
repeatable measurements. The base of the
5500A/COIL was designed so the current
clamp can be centered carefully on the coil,
minimizing operator error for best repeatabil-
ity. Calibration accuracy to specifications is
guaranteed only when proper clamp alignment
is made. The clamp current meter should be
centered as much as possible on the base dur-
ing calibration and verification. If the clamp-
type current meter has alignment marks, the
alignment marks should align the clamp with
the center bundled wire of the 50-Turn
5500A/COIL. (See Figure 1).
®
Fluke 5500a Coil Price
Figure 1. Positioning the Clamp